. 2005 Jan;43(1):468–471. doi: 10.1128/JCM.43.1.468-471.2005

TABLE 3.

Comparison of performance of VI with CSFV real-time RT-PCR assay

Test indexa % 95% confidence interval
Sensitivity 72.4 (50/69) 61.9-83.0
Specificity 100 (380/380) 100-100
Positive predictive value 100 (50/50) 100-100
Negative predictive value 95.2 (380/399) 93.1-97.3
False negative rate 27.5 (19/69) 17.0-31.9
False positive rate 0.0 (0/50) 0.0-0.0
Overall accuracy 95.7 (430/449) 93.9-97.6
a

Sensitivity, number of positive samples over number of true-positive samples; specificity, number of negative samples over number of true-negative samples; positive predictive value, number of samples that were positive and had CSFV; negative predictive value, number of samples that were negative and did not have CSFV; false-negative rate, number of negative samples confirmed positive over true-positive samples; false-positive rate, number of positive samples confirmed negative over true-positive samples; overall accuracy, total of true-positive and true-negative samples over the total number of samples.