Skip to main content
. 2005 Jan;12(1):224–227. doi: 10.1128/CDLI.12.1.224-227.2005

TABLE 2.

Precision and reproducibility of CD4 T-lymphocyte counts in a blood samplea

Sample no. Count
Dynabead Cyflow
1 252 344
2 292 341
3 254 351
4 278 347
5 264 354
6 258 356
7 300 356
8 296 344
9 280 359
10 316 360
    Mean 279 351
    SD 21.9 6.8
    CV% 7.8 1.9
a

Statistical comparison was done by a paired t test (P < 0.05).