Table 6.

Patient characteristics, including distinct electrocardiographic features, in relation to frequent false arrhythmia alarms

Patient clinical characteristics
Higher number of false arrhythmia alarms per monitored hour
Longer duration in seconds of false arrhythmia alarms per monitored hour
Clinical or electrocardiographic characteristic Median (range; sum) P-value Median (range; sum) P-value
Altered mental status (confusion or agitation documented) <0.001 <0.001
 Confused, n=198 0.5 (0–184; 800) 2.0 (0–2,263; 10,505)
 Not confused, n=263 0.0 (0–51; 381) 0.0 (0–3,812; 9,538)
^^Body mass index 0.794 0.456
 Body mass index ≥30, n=142 0.2 (0–51; 300) 0.5 (0–3,812; 7,939)
 Body mass index <30, n=314 0.7 (0–184; 870) 0.0 (0–2,263; 11,759)
Current smoker 0.327 0.516
 Smoker, n=71 0.0 (0–51; 144) 0.0 (0–3,812; 5,581)
 Nonsmoker, n=390 0.2 (0–184; 1,037) 0.5 (0–2,263; 14,463)
Seizures documented 0.614 0.769
 Seizures, n=19 0.2 (0–8; 30) 0.7 (0–184; 1,151)
 No seizures, n=442 0.1 (0–184; 1,151) 0.3 (0–3,812; 19,883)
Tremors documented 0.901 0.746
 Tremors, n=36 0.0 (0–26; 90) 0.0 (0–954; 1,408)
 No tremors, n=425 0.1 (0–184; 1,091) 0.4 (0–3,812; 18,653)
BBB per cardiac monitor 0.003 0.004
 BBB, n=41 0.6 (0–184; 339) 4.0 (0–3,812; 8,191)
 No BBB, n=420 0.0 (0–134; 842) 0.0 (0–2,263; 11,852)

Notes: Mann–Whitney U-test: number and duration of false arrhythmia alarms per hour compared between patients who do or do not have the selected characteristic (N=461);

^^

five patients had no recorded weight. Missing data was excluded from the calculation. Results in bold are statistically significant with alpha set at 0.05, patients with a characteristic depicted in bold have significantly more arrhythmia alarms than those patients without the characteristic.

Abbreviation: BBB, Bundle branch block.