Skip to main content
. 2017 Apr 24;7:37. doi: 10.1186/s13550-017-0272-5

Table 5.

Summary of published data on PET/CT parameters and patient effective dose for 18FDG whole body oncological studies

Reference Scan length as per this study Study from Study to Comment or CT description Tube current mAs or mA and s AEC CTDIvol (mGy) DLP (mGy cm) ED_CT (mSv) Activity (MBq) ED_PET (mSv)
[7] Torso Base of the skull Upper-thigh Impact of AEC 100–300 mA and 0.5 s Yes 7.2 370 6.23
250 mA and 0.5 s No 18.6
[36] Torso Base of the skull Upper-thigh TAI patients 112 mAs Yes 14.4 ± 2.8 312 ± 43 4.4 ± 0.6
[22] Torso Base of the skull Mid-thighs Standard CT 39 mAs Yes 438 ± 85 5.0 ± 1.0 450 ± 32 9.0 ± 1.6
Diagnostic CT 120 mAs 1050 ± 342 15.4 ± 5.0
[30] H&Torso Head Mid-thighs Initial CT Yes 6.4 ± 2.4 537 ± 222 8.1 ± 3.3
Optimized CT 4.3 ± 1.6 368 ± 145 5.5 ± 2.1
[20] H&Torso Top of the skull Mid-thighs Most frequent CT 110 mA and 0.8 s 4.3 434 7.5 336 ± 49 5.6
Special CT 250 mA and 0.5 s 6.0 616 10.95
Special CT 300 mA and 0.5 s 7.2 732 13.2
[8] H&Torso Vertex Mid-thighs 80 mAs Yes 8.2 304 ± 45 6.3 ± 1.2
[21] H&Torso Head Pelvis High-quality CT 80 mA and 0.8 s 19.0
High-speed CT 8.8
[9] H&Torso Head Mid-thighs Center 1 84 mA and 0.5 s Yes 5.9 ± 1 591 ± 130 7. 8 258 ± 56 4.9
Center 2 Yes 5.4 ± 2 575 ± 248 7.7 313 ± 73 5.9
WB Head Feet Center 1 89 mA & 0.5 s Yes 6.2 ± 1.4 1112 ± 285 9.4 258 ± 56 4.9
Center 2 Yes 3.7 ± 1 620 ± 166 5.9 313 ± 73 5.9
[23] Symphysis Low-dose CT 1–2.9 1.3–4.5 300–370 5.7–7
Contrast-enhanced CT 9.5–14.1 14.1–18.6
This study Torso Base of the skull Mid-thighs 120 mAs Yes 6.3 557 10.3 323 6.1 ± 1.2
HNT Vertex Mid-thighs 4.9 + 5.4 187 + 429 9.7 ± 2.2 299 ± 43 5.7 ± 0.8
WB Head Feet 7.9 + 3.3 776 + 307 14.0 ± 4.9 321 ± 71 6.1 ± 1.3