Skip to main content
. 2017 Feb 22;23(3):206–210. doi: 10.5152/dir.2016.15601

Table 2.

Fluoroscopy times, dose area products and skin doses for transjugular intrahepatic portosystemic shunts procedures in our cohort compared with reference levels from (22)

Range Mean±SD 75th percentile Median
Fluoroscopy time (min:s) Data set 04:48–112:40 12:09±14:38 26:08 16:17

RAD-IR DRL 60:00 31:00

DAP (Gy·cm2) Data set 5.1–479.6 40.3±73.1 75.2 38
RAD-IR DRL 525 252

Skin dose (mGy) Data set 37–3257 404±465 488.4 257
RAD-IR DRL 3000 1489

SD, standard deviation; RAD-IR, Radiation Doses in Interventional Radiology Procedures study; DRL, dose reference level; DAP, dose area product.