Skip to main content
. Author manuscript; available in PMC: 2018 May 1.
Published in final edited form as: ASAIO J. 2017 May-Jun;63(3):285–292. doi: 10.1097/MAT.0000000000000488

Figure 5.

Figure 5

PDF of the HMII VAD assembled to the inflow cannula at various circumferential orientations. (A) Single passage: 0° circumferential orientation generated the lowest SA main mode and the shortest high SA tail region (inset). (B) Repeated passages: significantly lower thrombogenic potential for the 0° circumferential orientation compared to the 30° and 60° orientations.