. 2016 Apr 29;7:11425. doi: 10.1038/ncomms11425

Table 1. Average irritant ranking score, comparison with negative control and degree of irritant for six samples.

Sample (four-week implantation) Average irritant ranking score Δ Between test sample and control Non-irritant (0.0–2.9)Slight irritant (3.0–8.9)Moderate irritant (9.0–15.0)Severe irritant (>15)
Negative control 4.0
Gel 18.0 14.0 Moderate irritant
Gel+IL 15.5 11.5 Moderate irritant
Gel+CNT 18.3 14.3 Moderate irritant
Gel+IL+CNT 15.3 11.3 Moderate irritant
Au 22.0 18.0 Severe irritant

CNT, carbon nanotube; gel, polyrotaxane-based gel with movable cross-linker (cyclodextrin/polyethylene); IL, ionic liquid; Negative control, high-density polyethylene sheet.

The assays were quantified based on ISO 10993-6:2007, Biological Evaluation of Medical Devices—Part 6: Tests for Local Effects After Implantation.