Table 1. Average irritant ranking score, comparison with negative control and degree of irritant for six samples.
Sample (four-week implantation) | Average irritant ranking score | Δ Between test sample and control | Non-irritant (0.0–2.9)Slight irritant (3.0–8.9)Moderate irritant (9.0–15.0)Severe irritant (>15) |
---|---|---|---|
Negative control | 4.0 | — | — |
Gel | 18.0 | 14.0 | Moderate irritant |
Gel+IL | 15.5 | 11.5 | Moderate irritant |
Gel+CNT | 18.3 | 14.3 | Moderate irritant |
Gel+IL+CNT | 15.3 | 11.3 | Moderate irritant |
Au | 22.0 | 18.0 | Severe irritant |
CNT, carbon nanotube; gel, polyrotaxane-based gel with movable cross-linker (cyclodextrin/polyethylene); IL, ionic liquid; Negative control, high-density polyethylene sheet.
The assays were quantified based on ISO 10993-6:2007, Biological Evaluation of Medical Devices—Part 6: Tests for Local Effects After Implantation.