Table 3.

Reportable quantities in EV analysis and relevant calibrators and standards

Measurement Derived from Calibrator or standards Reportable Values
EV concentration Number of events detected per volume of fluid Beads of known concentration to calibrate sample flow rate EVs/uL
EV size Signal intensity EV preps of known size (determined by orthogonal method, such as NTA or TRPS) or by calculation using standards of known size and refractive index. Diameter, surface area, or volume of detected EVs
EV fluorescence positive concentration Number of events detected above a threshold per volume of fluid Beads of known concentration to calibrate sample flow rate plus negative control sample to set threshold for positivity Positive EVs/uL above a threshold
EV fluorescence brightness/antibody number Signal intensity Standard particles with calibrated intensity or antibody capture capacity Mean fluorescence intensity (MESF) or antibodies bound per EV