Table 2.

Unadjusted Event Rates by PR Interval After Stratification by Device type

Outcome ICD CRT-D
PR<230 ms PR≥230 ms P Value PR<230 ms PR≥230 ms P Value
Heart failure hospitalization or death, n (%)
 1 y 1015 (15.8) 212 (18.8) 0.0108 2709 (16.9) 645 (22.2) <0.0001
 3 y 2083 (32.4) 441 (39.1) <0.0001 4900 (30.6) 1192 (41.0) <0.0001
 5 y 2540 (39.6) 530 (46.9) <0.0001 5727 (35.8) 1365 (47.0) <0.0001
Death, n (%)
 1 y 459 (7.1) 93 (8.2) 0.2034 1262 (7.9) 329 (11.3) <0.0001
 3 y 1322 (20.6) 29 1(25.8) 0.0003 3137 (19.6) 761 (26.2) <0.0001
 5 y 1787 (27.8) 398 (35.3) <0.0001 4009 (25.1) 957 (32.9) <0.0001
Heart failure hospitalization, n (%)
 1 y 702 (10.9) 151 (13.4) 0.0105 1903 (11.9) 432 (14.9) <0.0001
 3 y 1281 (19.9) 282 (25.0) 0.0002 3080 (19.3) 779 (26.8) <0.0001
 5 y 1513 (23.6) 329 (29.1) 0.0001 3466 (21.7) 866 (29.8) <0.0001

The log rank test was used for heart failure hospitalization or death and death end points; the Gray test was used for the heart failure hospitalization end point.

CRT-D indicates cardiac resynchronization therapy with defibrillator; and ICD, implantable cardioverter defibrillator.