Table 1. The measurement results (mean ± SD) under RT-setting and DR-setting.

Test Series Slice location Measurement index RT (mean ± SD) DR (mean ± SD) Mann-Whitney (P value)
Geometric accuracy (mm) T1 Localizer Phantom length 147.8±0.2 147.9±0.2 0.22
Slice 1 Diameter (AP direction) 190.7±0.3 190.7±0.2
Slice 1 Diameter (LR direction) 191.4±0.2 191.0±0.2
Slice 5 Diameter (AP direction) 190.7±0.5 190.6±0.3
Slice 5 Diameter (LR direction) 191.1±0.2 190.5±0.3
Slice 5 Diameter (phantom’s UL to LR) 190.8±0.3 190.6±0.2
Slice 5 Diameter (phantom’s UR to LL) 191.0±0.2 190.5±0.2
Mean diameter 191.0±0.4 190.7±0.3 <0.05
Slice thickness accuracy (mm) T1 Slice 1 Slice thickness 5.2±0.1 5.1±0.1 0.34
T2 Slice 1 Slice thickness 5.0±0.1 5.0±0.1
Slice position accuracy (mm) T1 Slice 1 Bar difference 1.4±0.7 0.7±1.0 0.41
Slice 11 Bar difference –2.2±0.4 –1.1±0.5
T2 Slice 1 Bar difference 1.5±0.2 0.8±0.9
Slice 11 Bar difference –2.2±0.2 –1.1±0.5
Image intensity uniformity (%) T1 Slice 7 PIU 93.4±0.1 92.7±0.2 <0.05
T2 Slice 7 PIU 93.3±0.2 93.1±0.3 <0.05
Percent-signal ghosting T1 Slice 7 Ghosting ratio 0.0005±0.0003 0.0006±0.0003 0.39
Low-contrast object detectability T1 Slices 8–11 Total number of spoke 31.0±1.3 36.9±0.8 <0.05
T2 Slices 8–11 Total number of spoke 23.8±1.7 32.1±1.6 <0.05

DR, diagnostic radiological; RT, radiotherapy; PIU, percent integral uniformity; LR, lower-right; UL, upper-left.