Table 2.

Experimental results of the ΔF+MA+GAUSSIAN.

Upstream Temp (°C) Downstream Temp (°C) Master Meter Flow Rate (m3/h) Master Meter Volume (L) Test Meter Volume (L) Volume Error (% of Read) Repeatability Error (%) Average Volume Error (% of Read)
38.96 39.41 17.69 591.90 590.40 −0.25 0.036 −0.27
38.97 39.41 17.81 591.40 590.00 −0.24
38.89 39.33 17.77 589.70 587.90 −0.31
38.80 39.17 35.80 1190.00 1189.50 −0.04 0.035 −0.07
38.72 39.16 35.84 1191.50 1190.80 −0.06
38.64 39.08 35.75 1188.20 1186.90 −0.11
38.56 39.00 88.42 2941.30 2947.50 0.21 0.077 0.20
38.55 38.99 88.55 2942.50 2950.70 0.28
38.48 38.91 88.55 2954.70 2958.40 0.13