. 2017 Apr 10;17(4):821. doi: 10.3390/s17040821

Table 3.

Experimental results of the DSC.

Upstream Temp (°C) Downstream Temp (°C) Master Flow Rate (m3/h) Master Total Volume (L) Test Total Volume (L) Volume Error (% of Read) Repeatability Error (%) Average Volume Error (% of Read)
29.21 29.37 88.95 1469.15 1473.66 0.31 0.07 0.24
29.21 29.37 88.89 1471.11 1475.04 0.27
29.21 29.29 89.09 1463.51 1465.70 0.15
29.13 29.29 37.42 619.59 619.47 −0.02 0.08 0.08
29.13 29.29 37.42 619.05 620.13 0.17
29.13 29.30 37.52 619.21 619.80 0.10
29.05 29.29 19.00 314.45 314.93 0.15 0.13 0.06
29.05 29.30 19.01 314.33 313.93 −0.13
29.04 29.21 19.00 314.40 314.91 0.16