Table 1. The recovery of FL-containing water samples detected by using the thin core-shell molecularly imprinted fluorescent nanosensor.

Samples Test FL added (nM) FL detected* (nM) Recovery (%)
MiniQ water 1 0 0.2 ± 0.03
2 30 27.7 ± 1.6 92.2 ± 5.6
3 60 57.5 ± 1.3 95.8 ± 2.3
4 120 123.3 ± 6.9 102.8 ± 5.6
5 500 425.1 ± 39.3 85.0 ± 9.3
Tap water 1 0 0.1 ± 0.04
2 30 34.7 ± 3.6 115.5 ± 10.3
3 60 62.7 ± 3.6 104.4 ± 5.7
4 120 124.6 ± 2.1 103.8 ± 1.6
5 500 405.7 ± 45.6 81.1 ± 11.2
Taihu Lake water 1 1 0 0.2 ± 0.04
2 30 33.3 ± 1.8 111.1 ± 5.3
3 60 63.3 ± 1.7 105.5 ± 2.7
4 120 128.6 ± 6.3 107.2 ± 4.7
5 500 400.3 ± 22.5 80.1 ± 5.5
Taihu Lake water 2 1 0 0.2 ± 0.1
2 30 34.2 ± 3.2 113.9 ± 9.4
3 60 61.5 ± 5.3 101.7 ± 8.7
4 120 128.6 ± 7.2 107.2 ± 0.9
5 500 417.8 ± 33.3 83.4 ± 7.9
Taihu Lake water 3 1 0 0.3 ± 0.1
2 30 33.1 ± 2.1 110.2 ± 6.3
3 60 62.3 ± 4.9 103.8 ± 7.8
4 120 124.9 ± 5.9 104.1 ± 3.9
5 500 410.2 ± 57.8 82.1 ± 14.1

*Average of three measurements.