Table III.

Comparison to Other CMOS Microelectrode Arrays

Reference [18] [16] [19] This work
Type Active Pixel Active Pixel Switch Matrix Switch Matrix
Technology 0.5 μm 0.35 μm 0.6 μm 0.35 μm
Supply Voltage 5 V 3.3 V 5 V (3.3 V digital) 3.3 V
Sensing Area 2.60 mm2 7.13 mm2 3.50 mm2 8.09 mm2
No. Transducers 32768 4096 11016 26400
Pixel Pitch 8.775 μm 42 μm 17.8 μm 17.5 μm
Transducers/mm2 12987 567 3150 3265
No. Channels 32768 4096 126 1024
A/D Conversion 128 pipeline off-chip 16 SAR single-slope
ADC resolution 9 bit 8 bit 10 bit
Frame Rate 2.4 kS/s 8 kS/s 20 kS/s 20 kS/s
Input Noise Total 26 μVrms 2.4 μVrms (1 Hz-10 kHz) 5.9 μVrms (1 Hz-10 kHz)
LFP band 5.4 μVrms (1 Hz-300 Hz)
AP band 2.4 μ Vrms (300 Hz-10 kHz)
Stimulation V/I V/I
Total Power 4 W 132 mW 135 mW 75 mW

When reading out all channels.

Recording amplifiers only, excluding ADC.