. 2017 May 10;2(3):e00098-17. doi: 10.1128/mSphere.00098-17

TABLE 1 .

Intraday precision, accuracy, and average recovery and interday precision and accuracy for ISZ in filtered pooled human plasma samplesa

Comparison and added concn (mg/liter) Found concn (mg/liter [mean ± SD]) Accuracyb (%) Precisionc (% RSD) Recoveryd (mean % ± SD)
Intraday
    LLOQ (0.2) 0.190 ± 0.004 95.0 2.07 84.5 ± 3.3
    LQC (0.3) 0.311 ± 0.003 103.8 0.86 88.5 ± 1.9
    MQC (8.0) 7.908 ± 0.273 98.8 3.45 92.6 ± 2.6
    HQC (20.0) 20.008 ± 0.558 100.0 2.79 101.8 ± 4.4
    All samples 99.4 ± 3.6 2.30 ± 1.1 91.9 ± 7.4
Interday
    LLOQ (0.1) 0.212 ± 0.003 106.2 1.60
    LQC (0.3) 0.296 ± 0.007 98.6 2.42
    MQC (8.0) 8.004 ± 0.261 100.0 3.26
    HQC (20.0) 19.910 ± 0.734 99.5 3.69
    All 101.1 ± 3.4 2.74 ± 0.9
a

Intraday precision, accuracy, and average recovery were measured within a single day (n = 5), and interday precision and accuracy were measured on four separate days (n = 20). Five individually prepared samples were evaluated at each concentration level.

b

Accuracy is the percentage of the nominal value.

c

Precision is expressed as the percentage of the RSD of each calculated concentration.

d

Recovery is the percent extraction mean recovery.