TABLE 3.

Stability of ISZ in plasmaa

Storage condition and sample concn (mg/liter) Found concn (mean mg/liter ± SD) Accuracyb (%) Precisionc (% RSD)
24 h at 20°C
    LQC (0.3) 0.300 ± 0.003 100.1 0.833
    HQC (20.0) 19.603 ± 0.142 98.0 0.727
7 days at 20°C
    LQC (0.3) 0.292 ± 0.005 97.2 1.652
    HQC (20.0) 19.373 ± 0.420 96.9 2.170
Freeze (−20°C) for 48 h
    LQC (0.3) 0.312 ± 0.008 104.0 2.606
    HQC (20.0) 20.265 ± 0.575 101.3 2.840
Freeze (−20°C) for 60 days
    LQC (0.3) 0.319 ± 0.040 106.4 12.545
    HQC (20.0) 22.119 ± 0.941 110.6 4.255
Three freeze-thaw cycles
    LQC (0.3) 0.308 ± 0.014 102.7 4.611
    HQC (20.0) 21.526 ± 0.500 107.6 2.321
Sample + IS for 24 h at 20°C
    LQC (0.3) 0.298 ± 0.009 99.2 3.082
    HQC (20.0) 21.654 ± 0.275 108.3 1.272
In autosampler for 24 h at 20°C
    LQC (0.3) 0.291 ± 0.005 97.0 1.598
    HQC (20.0) 19.172 ± 0.542 95.9 2.825
a

Intraday precision and accuracy values were determined for the LQC and HQC in filtered pooled human plasma samples after storage under various conditions. Five individually prepared samples were used for each concentration level and storage condition. IS, internal standard.

b

Accuracy is the percentage of the nominal value.

c

Precision is expressed as the percentage of the RSD of each calculated concentration.