. 2017 May 12;14:94. doi: 10.1186/s12985-017-0759-8

Table 4.

Patient samples that showed significantly different DNA quantification results using the Roche kit, the Daan kit, and the duplex real-time PCR assay

Sample no. Age Sex Cobas Daan Duplex-S Duplex-C
302-1 6 F 65.2 No 694 188
302-17 45 F 31.2 No 414 310.2
302-44 25 M 991 No 15010 15470
302-45 1 F 3.06E7 No 4.17E8 1.83E8
302-66 26 M 2.00E6 No 2.32E7 1.90E7
302-68 40 M 5.06E5 No 6.68E6 9.07E6
302-74 65 F 28.3 No 286 62.58
302-76 51 M 1.24E5 No 5.47E6 3.74E6
302-80 52 M 1.10E6 No 2.00E7 3.76E6
302-91 20 M 124 No 5437 488
302-102 45 M 81.9 No 2470 165
Chaoyang-9 62 F No 6.14E4 1.35E6 1.01E6
Chaoyang-36 82 M No 2.76E3 6.32E4 8.46E3
Chaoyang-53 26 F No 2.24E3 5.83E4 1.93E4

*Only patients whose HBV DNA quantitative results from the duplex real-time PCR were at least 1.0 log10 higher than that using Roche kit or Daan kit are listed. (Unit: IU/ml)