Skip to main content
. 2017 May 10;7:1699. doi: 10.1038/s41598-017-01841-x

Figure 1.

Figure 1

Schematic of multi-pass microscopy. A sample S is placed between two two objective and field lenses (OL and FL, respectively). This configuration is placed in between two mirrors M, which can be gated for in- and out-coupling of the electron beam (see methods). A pulsed probe beam is coupled into the optical path of the multi-pass microscope and illuminates S. The exit wave is subsequently re-imaged back onto the sample, which is now illuminated with an in-focus image of itself. This process is repeated multiple (m) times, after which the pulse is out-coupled and imaged onto a detector. For illustration, field (black) and imaging (red) rays are shown, which retrace themselves after one full roundtrip.