Figure 7.

Figure 7.

Low-resolution basket catheters are prone to false detections. A, Interelectrode distances in an 8-spline basket catheter. Resolution requirements for avoiding false detections for the 2 wavelengths (45.2 and 75.2 mm) are marked. B, Example isopotential map for longer wavelength simulation with basket electrodes marked. C, High-resolution phase map generated from phase at mesh vertices. Phase singularity (PS) marked as a black dot. D, Phase of electrodes as arranged on a regular 8×8 grid. E, Interpolated phase from basket arrangement of electrodes. F, Rotor PS locations computed from high-resolution data (blue) and the 8-spline basket electrodes (red) for the short wavelength simulation. Only PSs that correspond to the rotor are shown. G, All detected PSs of 8-spline basket catheter. H, PS detection histogram for 8-spline catheter. I, Rotor PS locations, (J) all PSs, and (K) PS detection histogram for a simulated 16-spline catheter—double the clinical resolution.