Table 3. Diagnostic accuracy parameters of 320-CTCA.

Classification Group Disease prevalence (%) Sensitivity (%) P value Specificity (%) P value Positive predictive value (%) P value Negative predictive value (%) P value Accuracy (%)
Per patient, n=107 Group 1, (HR≤60), n=55 69 97 (86–100) (37/38) >0.99 88 (64–99) (15/17) 1.00 95 (83–99) (37/39) 0.96 94 (70–100) (15/16) 0.37 96 (85–99) (52/55)
Group 2, (HR 60–80), n=52 67 100 (90–100) (34/34) 88 (64–99) (15/17) 95 (81–99) (34/36) 100 (78–100) (15/15) 96 (87–100) (50/52)
Per vessel, n=428 Group 1, (HR≤60) n=220 28 94 (84–98) (58/62) 0.92 91 (86–95) (144/158) 0.35 81 (69–89) (58/72) 0.32 97 (93–99) (144/148) 0.37 92 (87–95) (202/220)
Group 2, (HR 60–80), n=208 28 98 (91–100) (57/58) 94 (89–97) (137/146) 86 (76–94) (57/66) 99 (96–100) (137/138) 95 (91–98) (198/208)
Per segment, n=1,662 Group 1, (HR ≤60), n=855 13 89 (81–94) (95/107) 0.78 95 (93–97) (712/748) 0.54 74 (64–80) (95/128) 0.57 98 (97–99) (712/724) 0.28 94 (93–96) (807/855)
Group 2, (HR 60–80), n=807 13 84 (76–90) (90/107) 96 (94–97) (671/700) 76 (67–83) (90/119) 97 (96–99) (671/688) 94 (92–95) (761/807)

Diagnostic parameter values represented by parameter percentage and respective 95% confidence intervals with raw numbers for parameter calculation below.