Table 1.

Recorded peSPL values from the tubes and artificial ear simulator

Stimuli Small Diameter Tube Medium Diameter Tube Large Diameter Tube Ear Simulator

dBnHL peSPL peSPL peSPL peSPL
CLICK 85 118.2 117.0 114.1 116.5
88 121.1 120.0 117.0 119.4
91 124.1 122.9 120.0 122.3
94 126.9 125.8 122.8 125.1
97 129.8 128.5 125.6 127.8
100 132.2 131.0 128.1 130.2

500 Hz TB 85 102.8 101.1 97.4 108.4
88 105.9 103.8 100.5 111.5
91 109.0 106.8 103.2 114.7
94 112.3 110.2 106.5 118.0
97 115.5 113.3 109.7 120.9
100 118.5 116.3 112.0 123.8