Table 4.

Intra-assay and inter-assay accuracy and precision of DHA measurement in urine samples.

QC Level Concentration (ng/mL) Intraday (n = 6) Interday (n = 3)


%CVa %Biasb %CVa %Biasb
LLOQ 100 8.3 3.3 7.8 0.5
QC-low 300 5.7 −2.0 5.4 −4.6
QC-medium 800 4.7 6.5 7.4 3.1
QC-high 2000 4.5 8.5 3.4 4.4

Calibration range 100-5000 ng/mL, R2 = 0.9989 with weighing factor 1/X

QC, quality control; LLOQ, lower limit of quantification; CV, coefficient of variation.

a

% RSD = (standard deviation/mean) × 100.

b

%Bias = (mean calculated concentration – nominal concentration)/(nominal concentration) × 100.