Skip to main content
. Author manuscript; available in PMC: 2018 Feb 21.
Published in final edited form as: Phys Med Biol. 2016 Dec 20;62(4):1269–1290. doi: 10.1088/1361-6560/aa54c7

Figure 3. Pressure Calibration Comparison.

Figure 3

Plots show comparison of the acoustic pressures measured for each f-number as a function of driving voltage for direct pressure measurements (pDIR), summed sub-aperture pressure measurements (pSUM), and the estimated pressure levels predicted to account for the effects of nonlinear propagation (pEST). Pressure values reported for cavitation experiments were taken using pEST.