Biases between BioHarness and reference respiratory rate (Phase One: participants in sinus rhythm; Phase Two participants with atrial fibrillation)a.




Median br∙min-1 Bias br∙min-1 b Bias %
Phase One Rest REF 17.00 (6.75) -0.28 (4.00)tR -1.56 (23.42)
BH 16.15 (5.44)
Transition REF 19.30 (7.33) -2.20 (5.72)rR -12.17 (29.52)
BH 17.65 (6.32)
Run REF 41.70 (12.00) -1.36 (4.58)rt -3.30 (10.65)
BH 40.80 (10.09)
Total REF 39.90 (15.30) -1.25 (4.65)c -3.33 (12.01)
BH 39.02 (13.40)
Phase Two Rest REF 18.50 (6.65) -0.88 (4.30)twcsv -4.89 (21.77)
BH 17.29 (4.80)
Transition REF 19.70 (8.30) -5.73 (5.97)rwcsv -28.02 (23.69)
BH 14.34 (4.92)
Walk REF 22.30 (5.85) 0.81 (6.34)rtsv 3.12 (30.80)
BH 25.16 (6.35)
Cycle REF 25.00 (7.28) 0.28 (7.65)rtsv 1.04 (28.84)
BH 26.69 (6.94)
Sweep REF 22.00 (6.05) 6.61 (16.35)rtwcv 27.22 (77.73)
BH 31.03 (11.87)
Vacuum REF 19.65 (9.33) 9.42 (10.18)rtwcs 43.89 (67.69)
BH 31.55 (9.72)
Total REF 22.10 (7.18) 0.39 (7.33)c 1.56 (31.88)
BH 24.26 (11.02)

aTable reports median (IQR) reference (REF) and BioHarness (BH) respiratory rates, absolute (br·min-1) and relative (%) biases.

bThe letters r t R= rest, transition, run; statistically significantly different compared to Phase One rest, transition, Run (P<.001 to P=.04). The letters r t w c s v = rest, transition, walk, cycle, sweep, vacuum; statistically significantly different compared to Phase Two rest, transition, walk, cycle, sweep, and vacuum (P<.001 to P=.02).

cStatistically significantly different compared to reference measures (P<.001).