Table 3. .
Parameter |
Value or Description |
Class | 1M |
No. of lasers | 4 Superpulsed infrared |
Wavelength, nm, mean ± SD | 905 ± 1 |
Frequency, Hz | 250 |
Peak power, W each | 12.5 |
Average mean optical output, mW each | 0.3125 |
Power density, mW/cm2 each | 0.71 |
Energy density, J/cm2 each | 0.162 |
Dose, J each | 0.07125 |
Spot size of laser, cm2 each | 0.44 |
No. of red LEDs | 4 |
Wavelength of red LEDs, nm | 640 ± 10 |
Frequency, Hz | 2 |
Average optical output, mW each | 15 |
Power density, mW/cm2 each | 16.66 |
Energy density, J/cm2 each | 3.8 |
Dose, J each | 3.42 |
Spot size of red LED, cm2 each | 0.9 |
No. of infrared LEDs | 4 |
Wavelength of infrared LEDs, nm | 875 ± 10 |
Frequency, Hz | 16 |
Average optical output, mW each | 17.5 |
Power density, mW/cm2 each | 19.44 |
Energy density, J/cm2 each | 4.43 |
Dose, J each | 3.99 |
Spot size of LED, cm2 each | 0.9 |
Magnetic field, mT | 35 |
Irradiation time per site, s | 228 |
Total dose per site, J | 30 |
Total dose applied in muscular group, J | 180 |
Aperture of device, cm2 | 20 |
Application mode | Cluster probe held stationary in contact with skin at a 90° angle and using slight pressure |
Abbreviation: LED, light-emitting diode.