Table 3. Similarity indexes of lumen and stent strut detection algorithms.
Sensitivity | Specificity | Jaccard index | Dice index | |
---|---|---|---|---|
Lumen area | ||||
Auto vs. R1 | 98.69 ± 1.07% | 99.83 ± 0.15% | 97.38 ± 1.11% | 98.67 ± 0.58% |
Auto vs. R2 | 98.25 ± 1.30% | 99.76 ± 0.16% | 96.59 ± 1.19% | 98.26 ± 0.62% |
R1 vs. R2 | 98.76 ± 0.64% | 99.86 ± 0.01% | 97.62 ± 0.56% | 98.80 ± 0.29% |
Stent struts | ||||
Auto vs. R1 | 90.87 ± 9.44% | 94.75 ± 7.60% | 86.66 ± 10.08% | 92.53 ± 5.97% |
Auto vs. R2 | 91.27 ± 9.34% | 94.69 ± 7.54% | 87.00 ± 10.05% | 92.73 ± 6.00% |
R1 vs. R2 | 98.68 ± 3.73% | 98.49 ± 4.54% | 97.38 ± 5.28% | 98.6 ± 2.93% |
Auto—automatic detection algorithm; R1 –image reader 1; R2 –image reader 2