. 2016 Jul 7;9(7):543. doi: 10.3390/ma9070543

Table 1.

The sensitivity, linear range and detection limit of enzyme biosensors were tested on the first, seventh, 15th and 22nd days.

Days Analytical Characteristics NAF/LAC TiO2/LAC TiO2/NAF/LAC
1 Sensitivity (µA·L/µmol) 2.6 2.71 2.94
Linear Range/µM 1.25–150 1.25–150 0.75–150
Detection limit/µM 1.25 1.25 0.75
R2 0.9944 0.9986 0.9966
7 Sensitivity (µA·L/µmol) 2.78 2.62 2.85
Linear Range/µM 2.5–125 2.5–150 1.25–150
Detection limit/µM 2.5 2.5 1.25
R2 0.951 0.9962 0.9988
15 Sensitivity (µA·L/µmol) 2.6 2.4 2.8
Linear Range/µM 5–150 15–150 1.25–100
Detection limit/µM 5 3.75 1.25
R2 0.9993 0.9958 0.9966
22 Sensitivity (µA·L/µmol) 2.12 1.92 2.5
Linear Range/µM 5–150 3.75–150 3.75–150
Detection limit/µM 5 3.75 3.75
R2 0.9973 0.9988 0.9980