Skip to main content
. 2017 Jun 2;7:2714. doi: 10.1038/s41598-017-02821-x

Table 3.

Intra-day and inter-day precision (CV%) and accuracy of cetuximab quantification.

QC value (µg/mL) Intra-day Inter-day
Mean CV% Accuracy Mean CV% Accuracy
1 0.89* 8.0* 89.2* 0.933 14.7 93.3
1.8 1.66 9.5 92.0 1.74 9.7 96.8
90 79.7 2.8 88.6 88.4 9.5 98.3
175 178 2.5 101.7 180 12.4 103

Results from HT4 (A) and LT3 (B) peptides. *: Intra-day variability at LLOQ (1 µg/mL) was evaluated simultaneously with matrix effects in six individual batches of human plasma.