Summary of the hardware platforms currently available that could support closed-loop DBS

Platforms Advantages Disadvantages
Activa PC+S
  • Recording and stimulation on the same lead

  • System optimized for stimulation

  • Suitable for DBS-like use with much longer and frequent stimulation

  • Easily supports existing clinically effective approaches

  • Battery powered

  • The recording and stimulating systems are partitioned

  • Rapid battery depletion with frequent recording

Responsive Neurostimulation System (RNS)
  • Built-in signal processing algorithms

  • System configured by default for closed-loop control

  • System optimized for recording

  • Suitable for short and intermittent stimulation such as seizure-preventive stimulation, control of brief symptoms

  • Battery powered

  • Cannot record and stimulate at the same lead

  • Rapid battery depletion with frequent stimulation