Table 1. Analytical imprecision of three enzymatic techniques for measuring total bile acids.

Intra-assay Inter-assay
Mean (μmol/L) SD (μmol/L) CV Mean (μmol/L) SD (μmol/L) CV
Pool Low
Diazyme 16.60 0.07 0.44% 16.28 0.55 3.37%
Randox 14.68 0.16 1.12% 15.34 0.60 3.91%
Sentinel 18.35 0.31 1.66% 17.51 0.43 2.47%
Pool High
Diazyme 118.56 2.59 2.18% 118.69 2.40 2.03%
Randox 137.03 1.33 0.97% 146.03 3.47 2.37%
Sentinel 150.31 1.26 0.84% 149.65 1.52 1.02%