Table 1.

Results of Quality Grading for Vessel Depiction (summed up for all 15 locations, 0 = not present, 1 = partly/incomplete, 2 = clear/complete), Noise (mild = 0, moderate = 1, severe = 2) and Artifact impression (mild = 0, moderate = 1, severe = 2).

Vessel Depiction Noise Impression Artifact Impression
High-Venc Low-Venc Dual-Venc High-Venc Low-Venc Dual-Venc High-Venc Low-Venc Dual-Venc
Mean ± Std 20.2 ± 3.2 27.0 ± 1.6 27.2 ± 1.5 1.6 ± 1.5 0.5 ± 0.5 0.3 ± 0.3 0.8 ± 1 0.3 ± 0 0.1 ± 0
median 21 27 27 1.5 0.5 0.5 1 0 0

Plow_vs_dual 0.3750 0.0156 0.1875
Phigh_vs_dual < 0.0001 < 0.0001 < 0.0001
Phigh_vs_low < 0.0001 < 0.0001 0.0067