Table 1.

Suggested Protocol for Scanning to Total Knee Arthroplasty at 1.5T

Pulse Sequence
Parameters Axial Coronal Sagittal Sagittal Sagittal
Acquisition Type FSE FSE FSE MAVRIC-SL MAVRIC-SL
Weighting Intermediate Intermediate Intermediate Intermediate STIR
Repetition Time (ms) > 5000 > 5000 > 5000 4000–5000 4000–5000
Echo Time (ms) 30 30 30 40 40
Fat Suppression No No No No Inversion pulse at 150 ms
Echo Train Length 16–24 16–24 16–24 24 24
Receiver Bandwidth (Hz/pixel) 488.3 488.3 488.3 488.3 488.3
Flip Angle (deg) 90 90 90 90 90
Field of View (cm) 16 × 16 16 × 16 20 × 20 20 × 20 20 × 20
Acquisition Matrix 512 × 320 512 × 320 512 × 320 512 × 256 512 × 256
Slice Thickness (mm) 3 4 2.5 3.6 3.6
Section Gap (mm) 0 0 0 0 0
No. Signals Acquired 4 4 4 Variable Variable
In-plane Frequency Direction A to P R to L A to P S to I S to I
Acquisition Time 5–8 mins 5–8 mins 5–8 mins 5–7 mins 5–7 mins

Notes: A = Anterior, P = Posterior, R = Right, L = Left, S = Superior, I = Inferior