Figure 2.

Figure 2

Intracardiac electrograms during permanent His bundle lead threshold testing: With decreasing outputs during threshold testing of the permanent His bundle lead, loss of his capture and retrograde His activation are shown on the octopolar mapping catheter (curved red arrows and dashed arrows). This is accompanied by a subtle change in QRS morphology, but a clear increase in QRS duration (red circles).