Figure 7.

Figure 7

(a) Ion current measured as a function of the distance from the entrance of TW-SLIM escalator module. Error bars represent standard deviations from triplicate measurements. Measurements were made by inserting a conducting probe in between the two TW-SLIM surfaces at each point. The ~25% decrease at the entrance is attributed to m/z species not effectively transmitted by the SLIM (i.e., the mismatch between the ion injection and the SLIM ranges of optimal ion confinement and transmission). (b) Ion mobility resolution measured as a function of TW speed for a TW amplitude of 30 V, a TW sequence of 11110000, guard bias of 15 V, and rf amplitude (Vp-p) of 320 V for the linear TW-SLIM and TW-SLIM escalator arrangements.