Table 4. Within-run and between-run precision performance for variant frequency calling.
| Sample: variant | Within-run | Total | ||||||
|---|---|---|---|---|---|---|---|---|
| Replicates | Mean, % | SD, % | CV, % | Replicates | Mean, % | SD, % | CV, % | |
| YMC1: BRCA1, c.5496_5506delinsA | 3 | 45.27 | 1.03 | 2.27 | 5 | 45.72 | 1.28 | 2.79 |
| YMC2: BRCA2, c.2798_2799del | 3 | 47.53 | 0.15 | 0.32 | 5 | 49.44 | 2.61 | 5.29 |
| YMC3: BRCA1, c.922_924delinsT | 3 | 49.53 | 1.01 | 2.04 | 5 | 50.40 | 2.08 | 4.13 |
| YMC1: 8 heterozygote SNVs | 3 | 49.89 | 1.56 | 3.12 | 5 | 49.64 | 1.86 | 3.76 |
| YMC2: 7 heterozygote SNVs | 3 | 49.36 | 1.69 | 3.42 | 5 | 49.69 | 1.92 | 3.86 |
| YMC3: 14 heterozygote SNVs | 3 | 50.09 | 1.50 | 3.00 | 5 | 50.06 | 1.79 | 3.57 |
YMC = Yonsei Medical Center; SD = standard deviation; CV = coefficient of variation.