Skip to main content
. 2017 Jun;49(2):81–92.

Table 1.

Oxygenator operating and performance characteristics.

  SX18 FX15 Inspire6F
Manufacturer Terumo Cardiovascular Terumo Cardiovascular LivaNova
Surface area (m2) 1.8 1.5 1.4
Priming volume 270 144 284
Blood flow range (L/min) .5–7.0 .5–5.0 .5–6.0
Heat exchange surface area (m2) .22 .14 .43
Heat exchange material Stainless steel Stainless steel Polyurethane
Coating X Coating X Coating Phisio
Integrated filter surface area (cm2) 360 68
Integrated filter pore size (υ) 32 38
O2 transfer (mL/min) at 5 L/min—FiO2 100% 305 330 318
CO2 transfer (mL/min) at 5 L/min–1:1 gas-to-blood flow ratio 200 235 273
Pressure drop (mmHg) at 5 L/min 125 115 174

All values taken from manufacturers' IFU.