Skip to main content
. 2017 Jun;49(2):81–92.

Table 4.

Patient variables during CPB and oxygenator performance.

  SX18 FX15 Inspire6F p Value
Number 20 20 20
Blood flow (L/min) 4.3 ± .6 4.5 ± .6 4.5 ± .5 .302
SvO2 (%) 79.4 ± 5.8 82.5 ± 3.4 81.9 ± 5.4 .116
SVR 1,374 ± 340 1,350 ± 205 1,316 ± 339 .771
CPB mean line pressure (mmHg) 160.4 ± 20.9 209.0 ± 43.4 192.1 ± 24.6 .0001*
CPB mean sweep rate (L/min) 2.3 ± .6 2.4 ± .9 2.5 ± .5 .607
CPB mean FiO2 (%) 74.5 ± 6.54 79.1 ± 8.4 69.1 ± 7.9 .0005
CaO2 (mL O2/dL) 12.4 ± 1.1 13.6 ± 1.7 13.1 ± 1.4 .049
CvO2 (mL O2/dL) 9.5 ± 1.2 10.7 ± 1.5 10.3 ± 1.5 .021§
AVO2 difference (mL O2/dL) 2.9 ± .6 2.8 ± .5 2.8 ± .6 .800
DO2 (mL O2/min) 528.2 ± 102.3 611.1 ± 150.4 603.0 ± 80.6 .049
DO2/BSA (mL O2/min/m2) 291.1 ± 56.1 297.5 ± 38.7 288.4 ± 34.1 .799
VO2 (mL O2) 125.2 ± 32.6 127.4 ± 35.2 129.4 ± 31.1 .921
VO2/BSA (mL O2/min/m2) 67.9 ± 14.0 65.3 ± 12.1 65.7 ± 15.8 .810
O2 extraction .24 ± .07 .21 ± .03 .22 ± .05 .133

All data are mean ± SD.

*

SX18 vs. FX15 (p < .0001) and Inspire (p < .006).

Inspire vs. SX18 (p < .034) and FX15 (p < .0001).

SX18 vs. FX15 (p < .047).

§

SX18 vs. FX15 (p < .019).

SX18 vs. FX15 (p < .049).