Skip to main content
. 2017 Apr 27;8(5):2745–2755. doi: 10.1364/BOE.8.002745

Table 1. Corneal damage thresholds induced by 1338 nm laser at different spot sizes for slit lamp imaging at 6 hours post exposure.

Corneal spot diameter (mm) Number of eyes Number of exposures Damage threshold radiant exposure (95% fiducial limits) (J/cm2) Probit slope (ED84/ED50)
0.28 8 90 70.3 (68.6, 71.9) 1.40
0.94 12 89 35.6 (33.8, 36.3) 1.28
1.91 24 160 29.6 (28.3, 30.8) 1.14
3.55 12 48 30.3 (28.8, 31.8) 1.22