Skip to main content
. 2017 Mar 15;22(6):686–695. doi: 10.1177/2472555217696794

Table 1.

Acceptance criteria.

Parameter Acceptance Criterion
PPC well level
 RSEAL (baseline)a >100 MΩ
 Current amplitude (baseline) >0.1 nA
 RSEAL stability (between first and second additions) <50% decrease
PPC plate level
 Z′ factor ≥0.4
 Success rate (% valid wells) >90% accepted wells per PPC plate
 EC/IC50 for reference compounds ≤0.5 log from historical mean

PPC = population patch clamp.

a

Typical Rseal values ranged between 200 and 1000 MΩ.