. 2017 Jun 23;55(7):2180–2187. doi: 10.1128/JCM.00169-17

TABLE 3.

Overall specificity analysis for the Elecsys HTLV-I/II and Abbott Architect rHTLV-I/II assays in routine diagnostic samples

Parametera Elecsys HTLV-I/II serum Abbott Architect rHTLV-I/II serum
No. of samples
    Total 2,399 2,399
    Negative 2,336 2,336
    With ≥1 s/co
        IRpos 66 68
        IRfpos 4 7
        RRpos 66 68
        RRfpos 4 7
No. of samples with HTLV-1/2 immunoblot assay result/total no. tested
    Positive 59/66 59/68
    Indeterminate 3/66 2/68
    Negative 4/66 7/68
% specificity IR ≥ 1 s/co 99.83 99.70
95% confidence limit (two-sided, IR ≥ 1) 99.56–99.95 99.38–99.88
% specificity RR ≥ 1 s/co 99.83 99.70
95% confidence limit (two-sided, RR ≥ 1) 99.56–99.95 99.38–99.88
a

Total, number of analyzed samples; negative, number of true-negative samples (excluding indeterminate, positive, and false-negative samples); IR, initially reactive; IRpos, initially reactive sample, including true-positive samples; IRfpos, initially reactive sample, not including true-positive samples; RR, repeatedly reactive; RRpos, repeatedly reactive sample, including true-positive samples; RRfpos, repeatedly reactive sample, not including true-positive samples; s/co, signal/cutoff ratio.