Figure 2.

Figure 2

Renal replacement therapy (RRT) using RRT machine during extracorporeal membrane oxygenation (ECMO). If the ECMO circuit uses a roller pump, a proportion of the circuit blood comes from the patient via the venous drain cannula and enters the RRT machine where replacement, dialysis, and ultrafiltration occurs. Blood from the RRT machine then goes back to the ECMO bladder to the ECMO pump, the membrane oxygenator, and back to the patient via a return cannula. If a centrifugal pump is used, the RRT machine must be connected after the ECMO bladder to prevent air entrapment.