Table 2. The sensitivity and specificity of the four methods when analyzing field samples.

Detection target Method Sensitivity Specificity False-positive
rate
False-negative rate
O1 Culture 66.67% 100% 0 33.33%
Vch-UPT-LFa 100% 100% 0 0
Real-time fluorescent PCR 100% 94.05% 5.95% 0
Colloidal gold assay 91.67% 91.67% 8.33% 8.33%
O139 Culture 0b 100% 0 100%b
Vch-UPT-LF 100% 98.95% 1.05% 0
Real-time fluorescent PCR 100% 98.95% 1.05% 0
Colloidal gold assay 0b 97.89% 2.11% 100%b

a the up-converting phosphor technology-based lateral flow assay for the rapid and simultaneous detection of V. cholerae serogroups O1 and O139

b The low sensitivity (0) and the high-false negative rate (100%) for V. cholerae O139 for the culture method and colloidal gold assay occurred because there was only one positive sample for V. cholerae O139, while both of these methods yielded false-negative results.