Table 4.
Mean absolute relative difference (mean and median) values of the second generation Medtronic Paradigm Veo Enlite (ENL) sensor and the Dexcom SEVEN PLUS (7P), G4 PLATINUM (G4P), and advanced G4 for artificial pancreas studies (G4AP) sensors.
Sensor | Day of Analysis | MARD % | ||||
---|---|---|---|---|---|---|
Global | BG-IG | Calibration | Noise | |||
ENL | 2 | Mean | 11.7 | 3.6 | 11.3 | 4.2 |
Median | 9.8 | 3.3 | 10.3 | 3.2 | ||
7P | 2 | Mean | 14.2 | 3.5 | 12.8 | 5.6 |
7P | 1 | Median | 14.1 | 6.8 | 14.1 | 5.4 |
G4P | 1, 4, 7 | Median | 11.2 | 4.4 | 9.4 | 3.7 |
G4AP | 1, 4, 7 | Median | 10.0 | 3.4 | 9.4 | 3.7 |