Skip to main content
. 2017 Jun 12;17(6):1361. doi: 10.3390/s17061361

Table 4.

Mean absolute relative difference (mean and median) values of the second generation Medtronic Paradigm Veo Enlite (ENL) sensor and the Dexcom SEVEN PLUS (7P), G4 PLATINUM (G4P), and advanced G4 for artificial pancreas studies (G4AP) sensors.

Sensor Day of Analysis MARD %
Global BG-IG Calibration Noise
ENL 2 Mean 11.7 3.6 11.3 4.2
Median 9.8 3.3 10.3 3.2
7P A 2 Mean 14.2 3.5 12.8 5.6
7P B 1 Median 14.1 6.8 14.1 5.4
G4P B 1, 4, 7 Median 11.2 4.4 9.4 3.7
G4AP B 1, 4, 7 Median 10.0 3.4 9.4 3.7

A Values reported in [7]; B Values reported in [8]. Abbreviations: MARD, mean absolute relative difference; BG-IG, blood glucose to interstitium glucose.