Table 2.

Comparison of Agreement with Diagnostic Modalities Compared with Slit Lamp Examination

Testing Modality Mean Difference, % (SE) Limits of Agreement % Outside LOA P value*
Examiners −1.77 (−6.35 to 2.81) −22.44 to 18.89 4.55 0.31
AS-OCT −9.60 (−18.66 to −0.54) −50.47 to 31.27 13.64 0.008
Pentacam 10.35 (−3.63 to 24.32) −45.85 to 66.54 16.67 0.12
USP 21.83 (5.44 to 30.82) −38.93 to 82.49 5.56 0.07

Paired comparisons of % corneal thinning as measured by two masked graders with slit lamp examination, and mean grade by slit lamp compared with AS-OCT, Pentacam and USP. Abbreviations: SE, Standard Error of the measurement; LOA, Limits of Agreement; AS-OCT, Anterior Segment Optical Coherence Tomography; USP, Ultrasonic Pachymetry

*

Pitman’s test of difference in variance