Table 2.
Comparison of Agreement with Diagnostic Modalities Compared with Slit Lamp Examination
Testing Modality | Mean Difference, % (SE) | Limits of Agreement | % Outside LOA | P value* |
---|---|---|---|---|
Examiners | −1.77 (−6.35 to 2.81) | −22.44 to 18.89 | 4.55 | 0.31 |
AS-OCT | −9.60 (−18.66 to −0.54) | −50.47 to 31.27 | 13.64 | 0.008 |
Pentacam | 10.35 (−3.63 to 24.32) | −45.85 to 66.54 | 16.67 | 0.12 |
USP | 21.83 (5.44 to 30.82) | −38.93 to 82.49 | 5.56 | 0.07 |
Paired comparisons of % corneal thinning as measured by two masked graders with slit lamp examination, and mean grade by slit lamp compared with AS-OCT, Pentacam and USP. Abbreviations: SE, Standard Error of the measurement; LOA, Limits of Agreement; AS-OCT, Anterior Segment Optical Coherence Tomography; USP, Ultrasonic Pachymetry
Pitman’s test of difference in variance