Table 3.

Comparison of Agreement Between Diagnostic Modalities

Testing Modality Mean Difference, μm (SE) Limits of Agreement % Outside LOA P value*
AS-OCT - Pentacam 16.37 (−34.41 to 67.14) −206.72 to 239.45 4.76 0.01
AS-OCT - USP −76.42 (−128.45 to −24.40) −292.27 to 139.42 5.26 0.88
USP - Pentacam 97.77 (35.08 to 160.46) −162.37 to 357.91 5.26 0.18

Paired comparisons of remaining microns of cornea in area of maximal corneal thinning for AS-OCT, Pentacam, and USP. Abbreviations: SE, Standard Error of the measurement; LOA, Limits of Agreement; AS-OCT, Anterior Segment Optical Coherence Tomography; USP, Ultrasonic Pachymetry

*

Pitman’s test of difference in variance