Published in final edited form as: IEEE Trans Med Imaging. 2016 Apr 4;35(9):2098–2106. doi: 10.1109/TMI.2016.2550007

TABLE 1.

Shear Wave Speed Bias and CNR of Inclusion and Background

Inclusion (m/s) mean ± std. Inclusion Bias from nominal (%) Background (m/s) mean ± std. Background Bias from nominal (%) CNR (dB)
PROSE 16.7 mm; 400 Hz 4.85 ± 0.49 −6.08 2.60 ± 0.18 −11.3 22.1
10.4 mm; 400 Hz 4.40 ± 0.47 −14.8 2.50 ± 0.15 −13.4 22.1
10.4 mm; 350 Hz 4.26 ± 0.32 −17.5 2.56 ± 0.11 −11.3 23.8
10.4 mm; 300 Hz 3.63 ± 0.33 −29.7 2.59 ± 0.11 −10.3 19.5
6.5 mm; 400 Hz 3.70 ± 0.27 −28.3 2.55 ± 0.11 −11.7 20.4

LOGIQ E9 16.7 mm 4.86 ± 0.49 −5.85 25.5
10.4 mm 4.59 ± 0.42 −12.1 24.3
6.5 mm 4.17 ± 0.36 −19.3 22.0
Background 2.79 ± 0.11 −3.49