Skip to main content
. Author manuscript; available in PMC: 2018 Apr 1.
Published in final edited form as: IEEE Trans Biomed Circuits Syst. 2017 Jan 16;11(2):324–335. doi: 10.1109/TBCAS.2016.2609854

TABLE IV.

Comparison of this work with the state-of-the-art

Parameter This Work [37] [10] [18] [28] [38] [27] [14] [13]
Year 2016 2015 2015 2014 2014 2013 2013 2012 2007
Technology 0.18 μm 0.18 μm
HV
0.5 μm
HV
65 nm 0.35 μm
HV
0.18 μm
HV
65 nm 0.35 μm
HV
0.7 μm
HV
Supply Voltage (V) 3.3 18 12 3 15 12 ±2.5 20 +6/−9
Number of Channels 8 1 16 256 2 8 512 2 1
Max Istim (mA) 0.25 1.05 1.45 0.465 1 0.504 0.05 1 1
Resolution (ENOB) 8 (7.4) 7 (-) 9 (8) 5 (4.3) 5 (4.6) 6 (-) 5 (-) 5 (-) 7 (-)
Charge balance architecture biphasic + calibration H-bridge current mirror dynamic current mirror dynamic current mirror asymmetric H-bridge biphasic + calibration pulse insertion, offset regulation dynamic current mirror
Matching (%) 0.3 <0.01 3.6 ±2.5 0.5 0.1 0.05 2.24 N.A. 0.4
Stimulation Modes multipolar monopolar, bipolar multipolar multipolar monopolar bipolar monopolar monopolar monopolar