Table 5.

Dose measurements for μCT imaging systems reported in the peer-reviewed literature

CT instrument Dosimetry methods Imaging application Area Doses reported Ref.
BioScan nanoSPECT/CT EBT2 gafchromic films Multi-modality, oncology 8.3–3361 mGy [23]
CT Imaging TomoScope dual source scanner CTDI100 (ionization chamber)
LiF TLDs
Monte Carlo simulations
General 153–200 mGy [76]
Gamma Medica X-O (Not reported) Pulmonology 63.6 mGy [91]
Gamma Medica X-SPECT/CT (not reported) Multi-modality, oncology ~1000 mGy [56]
Gamma Medica X-SPECT/CT LiF TLDs Oncology 14–134 mGy [85]
GE CT120 CTDI100 (MOSFET) General 20.15–56.79 mGy [101]
GE CT120 Monte Carlo simulations General Whole body, mouse: 67–189 mGy
Whole body, rat: 42–119 mGy
[36]
GE eXplore Locus LiF TLDs General 126–229 mGy [87]
GE Locus Ultra Ionization chamber General 64–270 mGy [13]
GE Locus Ultra Ionization chamber Oncology 69–296 mGy [55]
GE RS—90 Ionization chamber Pulmonology Ungated: 120 mGy
Gated: 150 mGy
[78]
Imtek MicroCAT TLDs Oncology Organs: 210–760 mGy [86]
Imtek (Siemens) MicroCAT II LiF TLDs
ionization chamber
General Air: 78–97 mGy
Organ: 65–86 mGy
[81]
Imtek (Siemens) MicroCAT II NanoDot General 11–104.5 mGy [83]
Imtek (Siemens) MicroCAT II TLDs General Air: 55.6–61.5 mGy [84]
LaTheta LCT—200 TLDs Contrast agents 5–40 mGy [15]
Rigaku RmCT2 (PerkinElmer Quantum FX) FGDs General 16 mGy [79]
ScanCo vivaCT40 Ionization chamber Whole body scanning (mice) Polymer: 712.4 mGy
Air: 845.9 mGy
[43]
ScanCo vivaCT 40 CTDI (ionization chamber) Orthopedics Air: 939 mGy
Polymer: 441 mGy
[44]
Siemens Inveon TLDs Orthopedics 40–1257 mGy [42]
Siemens Inveon NanoDot General Organs: 9.4–163.1 mGy [83]
Skyscan 1076 (Not reported) Orthopedics 400 mGy [102]
Skyscan 1076 Ionization chamber
TLDs
Monte Carlo
Pulmonology Ionization chamber: 1640 mGy
TLDs, skin: 1104 mGy
TLDs, lung: 813 mGy
Monte Carlo, organs: 900 mGy–4.5 Gy
[29]
SkyScan 1076 Ionization chamber Orthopedics 166, 434 or 776 mGy [45]
Skyscan 1178 CTDI100 (ionization chamber)
LiF TLDs
General Organs: 295–507 mGy [77]
X-Strahl SARRP (Not reported) Radiotherapy planning 16, 32 or 64 mGy [103]
Custom dual energy system Not reported-data unpublished Oncology 260 mGy [57]