Table 5.
CT instrument | Dosimetry methods | Imaging application Area | Doses reported | Ref. |
---|---|---|---|---|
BioScan nanoSPECT/CT | EBT2 gafchromic films | Multi-modality, oncology | 8.3–3361 mGy | [23] |
CT Imaging TomoScope dual source scanner | CTDI100 (ionization chamber) LiF TLDs Monte Carlo simulations |
General | 153–200 mGy | [76] |
Gamma Medica X-O | (Not reported) | Pulmonology | 63.6 mGy | [91] |
Gamma Medica X-SPECT/CT | (not reported) | Multi-modality, oncology | ~1000 mGy | [56] |
Gamma Medica X-SPECT/CT | LiF TLDs | Oncology | 14–134 mGy | [85] |
GE CT120 | CTDI100 (MOSFET) | General | 20.15–56.79 mGy | [101] |
GE CT120 | Monte Carlo simulations | General | Whole body, mouse: 67–189 mGy Whole body, rat: 42–119 mGy |
[36] |
GE eXplore Locus | LiF TLDs | General | 126–229 mGy | [87] |
GE Locus Ultra | Ionization chamber | General | 64–270 mGy | [13] |
GE Locus Ultra | Ionization chamber | Oncology | 69–296 mGy | [55] |
GE RS—90 | Ionization chamber | Pulmonology | Ungated: 120 mGy Gated: 150 mGy |
[78] |
Imtek MicroCAT | TLDs | Oncology | Organs: 210–760 mGy | [86] |
Imtek (Siemens) MicroCAT II | LiF TLDs ionization chamber |
General | Air: 78–97 mGy Organ: 65–86 mGy |
[81] |
Imtek (Siemens) MicroCAT II | NanoDot | General | 11–104.5 mGy | [83] |
Imtek (Siemens) MicroCAT II | TLDs | General | Air: 55.6–61.5 mGy | [84] |
LaTheta LCT—200 | TLDs | Contrast agents | 5–40 mGy | [15] |
Rigaku RmCT2 (PerkinElmer Quantum FX) | FGDs | General | 16 mGy | [79] |
ScanCo vivaCT40 | Ionization chamber | Whole body scanning (mice) | Polymer: 712.4 mGy Air: 845.9 mGy |
[43] |
ScanCo vivaCT 40 | CTDI (ionization chamber) | Orthopedics | Air: 939 mGy Polymer: 441 mGy |
[44] |
Siemens Inveon | TLDs | Orthopedics | 40–1257 mGy | [42] |
Siemens Inveon | NanoDot | General | Organs: 9.4–163.1 mGy | [83] |
Skyscan 1076 | (Not reported) | Orthopedics | 400 mGy | [102] |
Skyscan 1076 | Ionization chamber TLDs Monte Carlo |
Pulmonology | Ionization chamber: 1640 mGy TLDs, skin: 1104 mGy TLDs, lung: 813 mGy Monte Carlo, organs: 900 mGy–4.5 Gy |
[29] |
SkyScan 1076 | Ionization chamber | Orthopedics | 166, 434 or 776 mGy | [45] |
Skyscan 1178 | CTDI100 (ionization chamber) LiF TLDs |
General | Organs: 295–507 mGy | [77] |
X-Strahl SARRP | (Not reported) | Radiotherapy planning | 16, 32 or 64 mGy | [103] |
Custom dual energy system | Not reported-data unpublished | Oncology | 260 mGy | [57] |