Table 1.

The Concordance correlation coefficients (CCC) and average percent difference between the MRF results from four different RF profiles and standard measurements, before and after corrections.

CCC 800TBW2 2000TBW8 4800TBW2 4800TBW8
T1 before 0.911 0.954 0.875 0.950
after 0.996 0.997 0.997 0.998
T2 before 0.755 0.978 0.704 0.961
after 0.986 0.998 0.952 0.974
Average Percent Difference 800TBW2 2000TBW8 4800TBW2 4800TBW8
T1 before −24.9% −15.6% −27.9% −14.3%
after −1.8% −2.7% −2.3% −0.7%
T2 before 43.1% 7.8% 47.6% 9.5%
after −8.8% −2.2% −14% −9.1%