. 2017 Jul 11;7:5117. doi: 10.1038/s41598-017-05383-0

Figure 2.

Figure 2

SNAP reduces the variability of mechanical measurements on the same gel in different locations. Elastic moduli of the same gel measured with three different instruments using the same kind of colloidal probes. The data were then analysed with different methods numbered from 1 to 4. In 1 to 3 ones, the deflection sensitivity was calibrated with a force curve on a stiff substrate and the force constant was determined from the analysis of the thermal with the AFM software (1); from the analysis of the thermal with the JAVA applet (2) and from the vibrometer measurements (3). In method 4, SNAP was applied: the force constant was measured using the vibrometer and deflection sensitivity was re-calibrated with the correction factor λ.